Home Precision Environment Aerosol Conditioning DIL 551 Static Dilution System
DIL 551 Aerosol Dilution System product image thumbnail
Topas GmbH Aerosol Dilution Systems

DIL 551 Static Dilution System

Manual static 1:10 aerosol dilution system for 28.3 L/min flow, providing reproducible particle-concentration reduction while retaining the aerosol size distribution.

Dilution factor
1:10
Aerosol flow
28.3 L/min
Setting
Manual
Counterpressure
±35 kPa
1:10 28.3 L/min Manual Static dilution
ISO 14644-3 VDI 2083 VDI 3491-6 VDI 3491-15

Model Overview

Model Configuration Primary use Selection note
DIL 551 1:10; 28.3 L/min Manual static 1:10 aerosol dilution system for 28 Select DIL 551 for manual 1:10 dilution at 28.3 L/min.

Key Features

Defined and reproducible dilution

Reduces aerosol particle concentration by a specified ratio while retaining the particle-size distribution of the sampled aerosol.

Capillary and HEPA-bypass principle

A small capillary flow retains the original particle concentration while particles are removed from the larger bypass flow before both streams are recombined.

Stable operation

Designed to provide a consistent dilution ratio during changing operating conditions in cleanroom, filter-test and aerosol-measurement workflows.

No auxiliary compressed-air supply

The passive flow arrangement uses the downstream instrument or process flow, avoiding a separate dilution-air supply for normal operation.

Overpressure and underpressure capability

Suitable for integration in measurement lines that operate under positive or negative pressure within the specified counterpressure limits.

Low-maintenance construction

Uses a compact fluidic layout with a HEPA-filtered bypass and requires only routine inspection and service of the flow path and filter elements.

Technical Specifications

Parameter Unit Value / Description
CONDITIONING
Purpose Dilution of aerosols
Sample condition, inlet Non-diluted aerosol
Sample condition, outlet Diluted aerosol
Dilution factor 10
Aerosol volumetric flow rate L/min 28.3
Setting method Manual
CONNECTIONS & OPERATING CONDITIONS
Power supply 12 V DC, 30 mA
Hose connector mm Ø 8, optional Ø 10
Pressure drop kPa 0.8–1.3
Environmental temperature °C 10–50
Maximum counterpressure kPa ±35
PHYSICAL
Dimensions (W × H × D) mm 110 × 160 × 300
Weight kg 1.5

Specifications follow the current Topas GmbH product information reviewed for this page. Confirm the final configuration, accessories and calibration documentation at quotation stage.

Applications

Verification of cleanroom test aerosols

Reduces high upstream challenge-aerosol concentrations to a range suitable for optical particle counters during cleanroom and clean-air-device testing.

Laminar-flow and safety-cabinet acceptance

Supports acceptance measurements on laminar-flow boxes, work benches, safety cabinets and smaller cleanroom installations.

Filter separation-efficiency measurement

Extends the usable concentration range of particle counters in upstream and downstream filter-efficiency measurements.

Highly concentrated aerosol analysis

Conditions concentrated aerosol samples before particle counting to reduce coincidence-related measurement errors.

Aerosol research

Provides a defined dilution stage for experimental aerosol generation, transport and particle-size-distribution studies.

Measurement-range extension

Allows particle-measurement instruments to analyse aerosol concentrations that would otherwise exceed their operating range.